Compensate Passive Probes

Each oscilloscope passive probe must be compensated to match the input
characteristics of the oscilloscope channel to which it is connected. A poorly
compensated probe can introduce significant measurement errors.
1 Input the Probe Comp signal (see
2 Press [Default Setup] to recall the default oscilloscope setup (see
3 Press [Auto Scale] to automatically configure the oscilloscope for the Probe
4 Press the channel key to which the probe is connected ([1], [2], etc.).
5 In the Channel Menu, press Probe.
6 In the Channel Probe Menu, press Probe Check; then, follow the instructions
32
Default Oscilloscope Setup"
Comp signal (see
"Use Auto Scale"
on-screen.
If necessary, use a nonmetallic tool (supplied with the probe) to adjust the
trimmer capacitor on the probe for the flattest pulse possible.
On the N2862/63/90 probes, the trimmer capacitor is the yellow adjustment on
the probe tip. On other probes, the trimmer capacitor is located on the probe
BNC connector.
"Input a Waveform"
on page 30).
on page 30).
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
on page 29).