Sweep speed
2 µs/div
5 µs/div
10 µs/div

Acquiring to Segmented Memory

When capturing multiple infrequent trigger events it is advantageous to divide the
oscilloscope's memory into segments. This lets you capture signal activity without
capturing long periods of signal inactivity.
Each segment is complete with all analog channel, digital channel (on MSO
models), and serial decode data.
When using segmented memory, use the Analyze Segments feature (see
Persistence with Segmented Memory"
across all acquired segments. See also
for details.
To acquire to
1 Set up a trigger condition. (See
segmented
memory
2 Press the [Acquire] key in the Waveform section of the front panel.
3 Press the Segmented softkey.
4 In the Segmented Memory Menu, press the Segmented softkey to enable
5 Press the # of Segs softkey and turn the Entry knob to select the number of
6 Press the [Run] or [Single] key.
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
1 µs/div
20 µs/div
for details.)
segmented memory acquisitions.
segments into which you would like to divide the oscilloscope's memory.
Memory can be divided into as few as two segments and as many as 250
segments.
Bits of resolution
8
9
10
11
12
on page 191) to show infinite persistence
"To set or clear persistence"
Chapter
10, "Triggers," starting on page 137
12
Acquisition Control
"Infinite
on page 127