Post-Repair Procedures - Agilent Technologies E8362B Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

Repair and Replacement Procedures

Post-Repair Procedures

Post-Repair Procedures
After the replacement of an assembly, you must perform the service procedures in the
order listed in
Table
Procedures referenced in this table are located in
Adjustments,"
unless specified otherwise.
Table 7-2
Related Service Procedures
Replaced Assembly
A1 keypad assembly
A2 display assembly
A3 front panel interface
board
A4 power supply
assembly
A6 SPAM board
A8 fractional-N
synthesizer board
A9 fractional-N
synthesizer board
A10 frequency reference
board
A11 phase lock board
A12 source 20
A13 frequency offset
receiver assembly
A14 system motherboard No adjustment needed
7-76
7-2.
Adjustments
and Other Procedures
No adjustment needed
No adjustment needed
No adjustment needed
No adjustment needed
No adjustment needed
LO Power Adjustment
Offset LO Adjustment
10 MHz Frequency
Reference Adjustment
Restore option data (Refer
to
"Repairing and
Recovering Option Data"
Appendix
B.)
No adjustment needed
3.8 GHz PMYO Frequency
Adjustment
Source Calibration
Adjustment
Offset LO Adjustment
PNA Series Microwave Network Analyzers
Chapter 3, "Tests and
Verification, Performance,
and Other Tests and Procedures
A1 Front Panel Keypad and RPG Test
Chapter 4
A2 Display Test
in
A1 Front Panel Keypad and RPG Test
and
A2 Display Test
None needed
Noise Floor Test
Trace Noise Test
System Crosstalk Test
Frequency Accuracy Test
The Operator's Check
The Operator's Check
Frequency Accuracy Test
in
Frequency Accuracy Test
Source Power Accuracy Test
Source Power Linearity Test
Frequency Accuracy Test
The Operator's Check
A1 Front Panel Keypad and RPG Test
and
A2 Display Test
The Operator's Check
Service Guide E8364-90026
E8362B, E8363B, E8364B
in
Chapter 4
in
Chapter 4
in
Chapter 4

Advertisement