3/4 Rack Module Top Replacement; Tools Required; Removal Procedure; Replacement Procedure - Agilent Technologies E4980A Service Manual

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

Replacement Procedure

3/4 Rack Module Top Replacement

3/4 Rack Module Top Replacement

Tools Required

Removal Procedure

Refer to Figure 6-1, Figure 6-2, Figure 6-3 and for this procedure.
Step 1. Disconnect the power cable from the E4980A.
Step 2. Remove the handle, front bumper and rear bumper. (item 1, 2, 3 )
NOTE
To remove the handle, rotate the handle to a vertical position and pull the arms outwards.
Step 3. Work the six TORX T20 screws (item 4) loose from the bottom.
Step 4. Remove the 3/4 Rack Module Top with the Wind Angle. (item 5)
Step 5. Remove the seven VENT BLOCKS (item 6) from the 3/4 Rack Module Bottom.
Step 6. Remove the two TORX T20 screws (item 7) fastening the Wind Angle.
Step 7. Remove the Wind Angle (item 8) from the 3/4 Rack Module Top.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
NOTE
When you fasten the Wind Angle with the two TORX T20 screws, set screwdriver's torque
to 21.1 in-lb.
When you fasten the 3/4 Rack Module Top with the six TORX T20 screws, set
screwdriver's torque to 21.1 in-lb.
90
Torque screwdriver, TORX T20
Chapter 6