Service
Module Core Replacement
Ordering a core replacement module
1. Contact Agilent (see www.agilent.com/find/assist).
2. Order a core replacement for your module (part number M9302-60003).
3. When the core replacement arrives, the package includes:
Entitlement Certificate
l
Replacement module
l
Calibration certificate for the replacement module
l
RMA number
l
Return instructions
l
Replacing the defective module
Before opening a packaged module for troubleshooting, ensure that all ESD (electrostatic discharge)
precautions are observed. Refer to
1. Power down the chassis.
2. Remove the defective module from the chassis.
3. Write down the serial number shown on the side shield of the defective module. You will assign this serial
number to the replacement module using the Agilent M9392A Serial Number Update Utility.
4. Remove the replacement module from the box and shipping material.
32
ESD
(page 10) for details.
Agilent M9302A PXI Local Oscillator Service Guide