Replacement Procedure

A2 Receiver Board Replacement

A2 Receiver Board Replacement

Tools Required

Removal Procedure

Refer to Figure 6-10 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 166
Step 2. Remove the front panel as described in "Front Panel Removal" on page 167.
Step 3. Remove two TORX T15 screws (item 1).
Step 4. Disconnect the all semi-rigid cables (item 2).
NOTE
The label on the semi-rigid cable shows the last two digits of the part number.
Step 5. Slide out the A2 receiver board (item 3).
Figure 6-10
A2 Receiver Removal
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
180
TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Open-end torque wrench, 5/16 inch (set to 10 in-lb); or 8mm with 9.5in-lbs can be
alternatively used (for reconnecting SMA connector)