PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B

Required Service Test Equipment

a
Critical Specifications
Equipment
Freq: 10 MHz to 20 GHz
Frequency counter
Accuracy : ±0.5 ppm
Min Freq: 1 MHz
Spectrum analyzer
Max Freq: > 4 GHz
Resolution BW: 300 Hz
Accuracy: ±0.0068 dB
Power meter
Power sensor
Freq: 10 MHz to 4.2 GHz
(All)
Range: –30 to +20 dBm
Power sensor
Freq: 50 MHz to 20 GHz
(E8362B)
Range: –30 to +20 dBm
Freq: 50 MHz to 40 or
Power sensor
50 GHz
(E8363B, E8364B)
Range: –30 to +20 dBm
Dynamic accuracy
None specified
test set
Compression
None specified
test set
Digital voltmeter
Resolution: 10 mV
Printer
N/A
d
N/A
Test software
a. Unless specified otherwise, equipment listed is required for all analyzer models.
b. P = Performance tests, A = Adjustments, T = Troubleshooting, V = System verification
c. If an accurate measurement of the dynamic accuracy specification is not required, the E4418A or
E4419A can be used.
d. The recommended model or part number for all equipment listed with a "P" in the Use column is
required for proper operation of this test software.
Microsoft® and Windows® are U.S. registered trademarks of Microsoft Corporation.
Service Guide E8364-90026
Recommended
Model or Part
Number
Test Instruments and Software
53151A
Opt 001
8565E
E4418B/19B
8482A
E4413A
8487A
Z5623A
Opt H01
Z5623A
Opt K01
Any
Any printer with Microsoft®
Windows® 2000 driver
N7840A
General Product Information
Required Service Test Equipment
Alternate
Model or Part
Use
Number
None
P, A,T
856xE
A,T
c
P, A,T
E4418A/19A
None
P, A,T
None
P, A,T
None
P, A,T
None
P
None
P
Any
T
P
None
P
b
2-9