Assembly Replacement Procedures
RF Section E4446A, E4447A, E4448A

Option 123 Assemblies

E4446A, E4447A, E4448A

Figure 11-31
Option 123 Assembly and Cable Locations

A34 Mixer

Removal
1. Drop the front frame. See
2. Refer to
3. Remove the three screws (1) that attach the A34 Mixer to the
352
Figure
11-31. Remove cables W93, W98, W99, and flat flex
cable W90.
standoffs.
page 314
for instructions.
Chapter 11