Agilent Technologies PSA Series Service Manual page 77

Spectrum analyzer
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

If the DC/AC Coupling modes are not properly functioning, a similar
table can be used.
Improper displayed signal amplitudes in some attenuator settings can
be isolated to either A13 or A14 based on which steps are incorrect.
Signal out of the attenuators can be measured by disconnecting the
appropriate semirigid cable and using a spectrum analyzer. Referring
to
20 dB and a 30 dB section. Verify that the attenuators are receiving the
correct switching sequences by using the following tables.
Table 3-1
A15 Settings
Attenuation
6 dB
(dB)
Select
J4 pin 9
0
High
6
Low
10
High
20
High
30
High
40
High
50
High
60
High
70
Low
Chapter 3
Coupling
DC Block Select
DC
AC
Figure
3-4, A14 has two 2 dB sections and A15 has a 6 db, a 10 dB, a
6 dB
10 dB
Bypass
Select
J4 pin 4
J4 pin 2
Low
High
High
High
Low
Low
Low
High
Low
High
Low
Low
Low
High
Low
Low
High
Low
Troubleshooting the RF Section (E4440A, E4443A, E4445A)
RF Section Description (E4440A, E4443A, E4445A)
J3 pin 9
High
Low
10 dB
20 dB
20 dB
Bypass
Select
Bypass
J4 pin 1
J4 pin 8
J4 pin 5
Low
High
Low
High
High
High
Low
Low
Low
High
High
High
Low
Low
High
Low
High
Low
DC Block Bypass
J3 pin 4
Low
High
30 dB
30 dB
Select
Bypass
J4 pin 7
J4 pin 6
Low
High
Low
Low
High
Low
Low
High
Low
High
High
Low
Low
Low
High
Low
Low
High
High
Low
High
High
Low
High
High
Low
High
77