Assembly Replacement Procedures

Front Frame

Front Frame

For most service situations, the front frame assembly can be "dropped"
from the deck without disconnecting any cables. Refer to the section

"Drop the Front Frame"

continue with the
Use ESD precautions when performing this replacement procedure.
CAUTION
Drop the Front Frame
1. Remove the instrument outer case. Refer to the
2. Refer to
3. Pull the front frame off of the deck until it is disengaged from the
4. At this point, the front frame can be placed flat on the bench for
314
"Removal"
Case"
removal procedure.
Figure
11-5. Using the T-10 driver, remove the 7 screws
(1) that attach the front frame assembly (2) to the deck.
disc drive.
service while still attached to the instrument. If you want to
completely remove the front frame, continue with the
section.
. To completely remove the front frame,
section.
"Instrument Outer
"Removal"