Post-Repair Procedures

Configuring a Replacement CPU Assembly

Configuring a Replacement CPU Assembly

Description

When only the main CPU board is replaced, it is necessary to
reconfigure it with certain data that is critical for instrument operation.
This information includes the IP address, time, date, instrument serial
number, and the host name. If possible, it is best to capture this
information from the instrument before replacing the main CPU board.
A PC and the PSA Series Spectrum Analyzers Test and Adjustment
NOTE
software is required to restore the serial number.

Procedure

Capturing Critical Data for the CPU Board
Ensure that all information is recorded exactly as it is displayed.
1. To view the host name and IP address, press
The Ethernet Number is embedded in the CPU, and it will change
NOTE
when the CPU is replaced. When the Ethernet Address changes the
Flash Memory board will rest the IP address and host name to the
default factory values.
2. To view the instrument serial number, press
422
.
Configure I/O
(2 of 3) is displayed, and
proceeding.
a. Confirm that the serial number on screen matches the number on
the rear panel tag.
b. If the instrument is not functional, record the serial number on
the rear panel tag.
System
System
. Refer to the note below before
Show System
and
,
until
More
Chapter 12