Assembly Replacement Procedures
RF Section E4440A, E4443A, E4445A
Figure 11-10
RF Section Hardware

A18 YTO and A19 RYTHM

Removal
1. Refer to
2. Remove the ribbon cables attached to the YTO and RYTHM.
3. Refer to
4. Carefully remove the bracket containing the YTO and RYTHM from
322
Figure
11-9. Remove the semi-rigid cables, W9, W35, W36,
and W38.
Figure
11-10. Using the T-10 driver, remove the 2 screws
(1). Remove the third screw that can be accessed down behind the
YTO, near the Mid Web.
the RF section.
Chapter 11