Post-Repair Procedures

28 V Rear Panel Output Check

28 V Rear Panel Output Check
Description
Instruments with Option 219 and instruments with Front End Driver
assemblies E4440-60242 (<27 GHz models) or E4446-60059 (> 27 GHz
models) have a switchable 28 V connector on the rear panel.
Connect a voltmeter to the rear panel Noise Source Drive Out
connector. To turn on and off the 28 V, press
then
. Key in
System
Service
−49
the password
, and press
. Press
then press the
Enter
Service
Noise
key to toggle the 28 V on and off. The voltage should read 28 V
Source
±0.2 V.