Assembly Replacement Procedures

Instrument Outer Case

Instrument Outer Case
If the instrument is placed on its face during any of the following
CAUTION
procedures, be sure to use a soft surface or soft cloth to avoid damage to
the front panel, keys, or input connector.

Removal

1. Disconnect the instrument from ac power.
2. There are two handles on the sides of the instrument that must be
3. Remove the four bottom feet (2). This is done by lifting up on the
4. Remove the four screws (3) that hold the rear feet (4) in place.
5. Pull the instrument cover (5) off towards the rear of the instrument.
Figure 11-3
Instrument Outer Case Removal
310
taken off. Refer to
Figure
screws that attach each handle (1). Remove the handles.
tabs on the feet, and sliding the feet in the direction indicated by the
arrows.
11-3. Using the T-20 driver, loosen the
Chapter 11