Assembly Replacement Procedures
RF Section E4446A, E4447A, E4448A
Replacement
1. Position the RF section in the slots in the deck. Gently slide it down,
2. Using the T-10 driver, replace the 4 screws. Torque to 9 inch pounds.
3. Replace the cables to the correct locations. Torque the semi-rigid
4. Replace the cable hold-down. Torque the single screw to 10 inch
5. Refer to
Figure 11-22
RF Section Ribbon Cable Locations - E4446A, E4447A, E4448A
338
making sure that no cables get caught.
cables to 10 inch pounds including the W9 cable at the input
attenuator.
pounds.
Figure
11-22. Replace the ribbon cables to the correct
locations.
Chapter 11