Figure 11-33 SW3 and Bracket
Replacement
1. Attach SW3 to the bracket using the two screws removed earlier.
Torque to 9 in-lbs.
2. Place the switch/bracket assembly into place on the main bracket.
Refer to
SW3 bracket to the main bracket. Torque to 9 inch-pounds.
3. Refer to
9 in-lbs.
4. Replace ribbon cable W96.
5. Replace the front frame.

SW5 Switch

Removal
1. Drop the front frame. See
2. Refer to
cable W92.
3. Remove the three screws (2) that attach SW5 to the bracket. Lift
the switch/bracket assembly from the instrument.
4. Refer to
remove the three screws.
Chapter 11
Figure
11-31. Replace the two screws (4) that attach the
Figure
11-32. Replace cables W89. W93, and W94. Torque to
page 314
Figure
11-31. Remove cables W99, W100, W105 and ribbon
Figure
11-34. To separate the switch from the bracket,
Assembly Replacement Procedures
RF Section E4446A, E4447A, E4448A
for instructions.