Agilent Technologies PSA Series Service Manual page 328

Spectrum analyzer
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Assembly Replacement Procedures
RF Section E4440A, E4443A, E4445A
A36 Preamplifier and SW6
To remove either the A36 Preamplifier or SW6 switch, it will be
necessary to drop the front frame. To drop the front frame with Option
110 installed, follow these steps:
1. Refer to
2. Refer to
3. Drop the front frame. See
Figure 11-14
Option 110 A36 Preamplifier and Switch SW6
Option 110 SW6 Switch
Removal and Replacement
1. Refer to
2. Remove the two screws (2).
3. To replace SW6, install it onto the bracket and secure it with the two screws
4. Replace W112 and W113. Torque to 10 inch-pounds.
328
Figure
11-12. Remove cables W109 and W110.
Figure
11-13. Remove ribbon cable W114 and wire harness
W106 from the Option driver board.
Figure
11-14. To remove SW6, remove cables W112 and W113.
(2) removed earlier. Torque to 9 inch-pounds.
page 314
for instructions.
Chapter 11

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4440aE4443aE4445aE4446aE4447aE4448a

Table of Contents