Assembly Replacement Procedures
Front Frame Subassemblies

Bezel and Keypad

Removal
1. Remove the display/rubber mount assembly. Refer to the first section
2. Remove the RPG knob and volume knob by pulling straight off of the
3. Refer to
4. Refer to
5. Lift the front frame off of the keypad/keyboard assembly.
6. The bezel and keyboard can now be separated by pressing on the pin
Take care to not touch the contacts on the keypads. Contaminants on
NOTE
the contacts might interfere with the performance of the key.
Figure 11-62
Keypad Removal
404
of the
"Display Removal"
control shafts.
Figure
11-62. Using the T-10 driver, remove the 10 screws
(1) that secure the bezel and keyboard assembly to the front frame.
Figure
11-63. The 10 screws also attach the dress panel,
bezel, subpanel, keypad, and keyboard to the front frame. Take care
to keep these parts in the correct order and aligned properly.
just above the
On/Standby
unlock the tabs. The flexible keypads can be separated from the
keyboard and the bezel by pulling them apart.
procedure.
LEDs on the bezel and sliding apart to
Chapter 11