RPG

Removal/Replacement
1. Remove the front panel interface board. Refer to the
2. Unsolder the RPG, remove the nut and washer, and remove it from
3. Place the new RPG in the correct position, tighten the nut and
4. Replace the front panel interface board. Refer to the

External Trigger Cable

Removal/Replacement
The front panel External Trigger connector/cable must be replaced as
NOTE
an assembly.
Be careful to not scratch the dress panel when removing or replacing
CAUTION
this part.
1. Remove the front frame assembly. Refer to the
2. Refer to
3. Disconnect the cable from the A8 analog IF assembly.
4. For replacement, position the connector/cable through the front
5. Using the 9/16 socket, replace the nut to secure the connector to the
6. Clip the cable into the cable clamps on the shield.
7. Re-route the cable to avoid interference with the fans or the airflow,
8. Replace the front frame. Refer to the
Chapter 11
Interface Board"
removal procedure.
the front panel interface board.
washer, and resolder the leads.
Interface Board"
replacement procedure.
removal procedure on
Figure
11-64. Using the 9/16 socket, remove the nut (1)
that secures the connector to the front frame.
frame, matching the "D" slot.
front frame. Torque to 21 inch pounds.
and reconnect to the A8 analog IF assembly.
procedure.
page
314.
"Front Frame"
Assembly Replacement Procedures
Front Frame Subassemblies
"A2 Front Panel
"A2 Front Panel
"Front Frame"
replacement
407