Agilent Technologies PSA Series Service Manual page 424

Spectrum analyzer
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Post-Repair Procedures
Configuring a Replacement CPU Assembly
Entering the Serial Number
1. Load the PSA Series Spectrum Analyzers Test and Adjustment
2. Under Test Plan, select Utilities.
3. Under Available Tests, choose Serial Number Initialization.
You have only one chance to enter the serial number into the
CAUTION
replacement processor memory. Ensure that the number you enter
matches the number on the serial number tag on the rear panel. Use
extreme care when typing the serial number.
4. Follow the prompts and enter the serial number.
Entering the Time and Date
1. Press
2. Press
3. Press
Entering the IP Address and Host Name
To enter the IP Address and Host Name, refer to
IP Address and Host Name" on page
424
software on your PC.
,
System
Time/Date
, and enter the time in the format hh.mm.ss. For
Set Time
example, 15.25.30 for 15 hours 25 minutes and 30 seconds. Press the
softkey.
Enter
and enter the date in the format yyyymmdd. For
Set Date
example 20020311 for May 11, 2002. Press the
and select the Date Format.
427.
softkey.
Enter
"Step 5: Entering the
Chapter 12