Replacement - Agilent Technologies PSA Series Service Manual

Spectrum analyzer
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Replacement

1. Carefully position the top brace on the deck. There is an alignment
pin on the mid web/fan assembly that should match up with the
alignment hole on the top brace. Make sure that no coaxial cables
will get pinched underneath the brace.
2. Using the T-10 driver, replace the top screws first, and then the side
screws after the top screws are tightened. Torque to 9 inch pounds.
3. Replace the outer case. Refer to the
replacement procedure.
Chapter 11
Assembly Replacement Procedures
"Instrument Outer Case"
Top Brace
313