Front Frame Subassemblies - Agilent Technologies PSA Series Service Manual

Spectrum analyzer
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Front Frame Subassemblies

In order to remove any of the following subassemblies, it is necessary to
drop the front frame assembly from the main deck. Refer to the
the Front Frame"
front shield from the front frame assembly. Refer to
Remove the cables from the cable clamps. Using the T-10 driver, remove
the 11 screws (1) securing the front shield (2) to the front frame. Lift
the shield from the front frame.
After the subassembly is changed, replace the front shield using the 11
screws removed earlier. For alignment purposes, tighten the screws
marked with #1 and #2 first (silkscreened on the shield), then tighten
the remaining screws. Torque to 9 inch pounds. Reconnect the front
frame to the deck. Refer to the
Use ESD precautions when performing the following replacement
CAUTION
procedures.
Figure 11-55
Front Frame Shield
Chapter 11
procedure. It will also be necessary to remove the
"Front Frame"
Assembly Replacement Procedures

Front Frame Subassemblies

"Drop
Figure
11-55.
replacement procedure.
395