Hardware Options
Verifying Option 124, Y-Axis Video Out
6. Place the displayed signal at mid screen by pressing
7. Adjust the reference level to place the signal exactly on the top
The signal trace cannot be displayed above the top graticule line.
NOTE
However, the analyzer will measure signals above the top graticule line.
Therefore the Video Output will be driven above 1 V even though it
appears the on screen trace is only at the top graticule line.
8. Adjust the reference level to place the signal on the bottom graticule
9. Notice also that as the signal is moved up and down the screen, the

Procedure 2 - Detailed view of the video signal

Connect channel 1 of the oscilloscope to the rear panel Video Out port of
the PSA.
Connect channel 2 of the oscilloscope to the rear panel Trigger 1 Out
port of the PSA. Trigger 1 on the PSA is the High=Sweeping (HSWP)
signal. Trigger 1 goes high just before the sweep starts on the analyzer
screen. These connections will allow the oscilloscope display to resemble
the PSA display.
PSA Setup:
Instrument preset
Turn on internal 50 MHz,
−25 dBm cal signal
Center Frequency
Span
Sweep Time
Res BW
Attenuator
Ref Level
Scale/Div
204
adjusting the reference level until the signal is as close as possible to
exactly mid screen. The voltmeter should read 0.5 volts.
graticule line. The voltmeter should read 1 V.
line. The voltmeter should read 0V.
voltmeter reading changes 0.1V per graticule division.
50 MHz
10 MHz
1 ms
300 kHz
10 dB
−24 dBm
Places the 50 MHz cal signal
at top screen on the PSA
10 dB/ Div
, and
Amplitude
Chapter 8