Troubleshooting the Processor, Power Supply, and Display

Isolating an A5 Power Supply Problem

Isolating an A5 Power Supply Problem
There are no fuses to replace within the power supply. If you determine
NOTE
that the power supply is the failed assembly, replace the power supply.
Observing the LED on the front of the instrument, and measuring the
probe power connector, will determine if there is catastrophic failure in
the power supply assembly.
1. Ensure the instrument is plugged in with the power switch in the
2. Power on the instrument and verify that the green LED on the front
3. The front panel probe power connector can be used to check the
Figure 7-1
Probe Power Connector
If all of these supplies seem dead, it is likely that the problem is a
defective power supply assembly, or some other assembly is loading
down the power supply. Continue with
Dead" on page
If the correct LEDs are lit and the probe power voltages measure within
the specifications, the power supply has not suffered a catastrophic
failure; however, the power supply could still be at fault. Continue with
the next section to measure the individual voltage supplies.
154
Standby position (power not switched on). Verify that the yellow
LED next to the power switch is lit. A lit yellow LED indicates the
+15 VDC line (P15 STB) is providing enough voltage to light the
LED. (The actual voltage may not be +15 VDC.)
panel is lit. A lit green LED indicates the power supply has received
an "ON" command and that the +5.2 VDC supply can at least light
the LED.
+15 VDC and −12.5 VDC (−15 VDC) supplies. The −12.5 VDC is
produced by post regulating the −15 VDC supply. Refer to
for a diagram of the probe power connector.
155, to determine the cause of the problem.