Assembly Replacement Procedures
A5 Power Supply

Replacement

1. Place the power supply assembly into position in the deck. Push
down to mate the assembly with the motherboard connector.
2. Refer to
Figure
11-37. Using the T-10 driver, replace the 5 screws.
Torque to 9 inch pounds.
3. Replace the instrument top brace. Refer to the
replacement procedure.
362
"Top Brace"
Chapter 11