Agilent Technologies PSA Series Service Manual page 400

Spectrum analyzer
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Assembly Replacement Procedures
Front Frame Subassemblies
Figure 11-59
Front Frame Subassemblies
Display Replacement
1. Carefully place the display into the rubber display mount. Flex the
2. Refer to
3. Place the display mount / display into the front frame.
4. Route the bottom display backlight cable from the display under the
5. Plug the bottom backlight cable from the display into the backlight
6. Plug the top backlight cable from the display into the top inverter
7. Plug the Display Converter board/flat flex cable into the flat panel
8. Place the ferrite block (on the flat flex cable) into the rubber display
9. Press the tape down on the flat flex cable to secure the cable to the
400
display mount rubber tabs until they are in place over the edges of
the display.
Figure
11-59. Install the pressure plate (8) between the
display mount and the top of the display.
rubber tab of the display mount.
extension cable (4). Press the backlight extension cable into the
grooves of the display mount.
board. Press the backlight extension cable into the grooves of the
display mount.
display connector.
mount.
display.
Chapter 11