Table 1-3
Required Test Equipment for PSA Series (Continued)
Instrument
Millimeter Power Splitter
(for E4446A, 47A, 48A)
Directional Bridge
Directional Coupler
DC Probe
High Frequency Probe
Negative Detector
Bias Adjustment Board
Cables
3.5 mm (m) to 3.5 mm (m)
(2 required)
2.4 mm (f) to 2.4 mm (m)
(for E4446A, 47A, 48A)
Type-N
(2 required)
Chapter 1
Critical Specifications
Frequency: 10 MHz to 50 GHz
VSWR: ≤ 10 MHz to 18 GHz: < 1.29:1
18 GHz to 26.5 GHz: < 1.20:1
26.5 to 40 GHz: < 1.50:1
40 GHz to 50 GHz: < 1.65:1
Tracking Error: < ± 0.40 dB
Connector: 2.4 mm (f)
Frequency Range: 5 MHz to 3 GHz
Directivity: ≤ 5 MHz: 30 dB
5 MHz to 2 GHz:40 dB
2 GHz to 3 GHz: 30 dB
VSWR: ≤ 2 GHz: ≤ 1.15:1
≤ 3 GHz: ≤ 1.22:1
Insertion Loss: ≤ 1.5, +0.1 dB/GHz
Coupling (nominal): 16 dB
Connector: Type N (f)
2 GHz to 20 GHz
Directivity > 16 dB
Transmission arm loss: < 1.5 dB (nominal)
Coupled Arm Loss: ~10 dB (nominal)
VSWR: ≤ 1.35:1
Connector: SMA (f)
Frequency Range: 300 kHz to 3 GHz
Input Resistance: 1 M Ω (nominal)
50 MHz to 26.5 GHz
± 0.6 dB to 18 GHz
Frequency: DC to 26.5 GHz
Length: ≤ 92 cm (36 in)
Insertion Loss: ~2 dB
VSWR: DC to 18 GHz: ≤ 1.25:1
18 GHz to 26.5 GHz: ≤ 1.35:1
Frequency: DC to 50 GHz
Length: ≤ 24.9 cm (9.8 in)
Insertion Loss: ≤ 26 GHz, ~4 dB
Insertion Loss: ≤ 40 GHz, ~5 dB
Insertion Loss: ≤ 50 GHz, ~6 dB
VSWR: ≤ 26.5 GHz: ≤ 1.30:1
≤ 40 GHz: ≤ 1.40:1
≤ 50 GHz: ≤ 1.55:1
Frequency: 10 MHz to 8 GHz
Precision Type-N (m), both ends
62 cm (24 in.)
VSWR: ≤ 18 GHz: 1.4:1
Insertion Loss: 1.5 dB
Service Equipment You Will Need
Recommended
Alternative
Agilent Model
Agilent Model
Number
Number
11667C
86205A
87300B
11002A
11003A
85024A
33330C
E4440-60041
8120-4921
8120-6164
11500C
Overview
a
Use
P
P
P
A
T
A
A
A, P
A, P
A, P, T
29