Display Filter Removal/Replacement
1. Remove the rubber mount containing the display assembly from the
Be careful when handling the display/mount assembly outside of the
CAUTION
front frame. The individual components are not secured in the mount
and could possibly fall out of the mount unless it is held together.
2. Flip the display/mount assembly over and place it filter-side up on a
3. The display filter can now be removed by pulling back on one of the
4. When you install a new filter, take care to touch only the outside
Make sure the side of the filter with the silver edge is facing up, away
NOTE
from the display.
5. The display/mount assembly can now be reinstalled into the front
Figure 11-60
Display Filter Replacement
Chapter 11
front frame. Refer to the
flat surface.
corner supports (2) and lifting the filter (1) out of the mount as
shown in
Figure
11-60.
edges. Install the filter into the mount by sliding under the corner
supports.
frame. Refer to the
"Display Replacement"
Assembly Replacement Procedures
"Display Removal"
procedure on
procedure on
Front Frame Subassemblies
page
399.
page
400.
401