Top Switch Replacement
The replacement switch will not have the metal EMI shield installed.
NOTE
The shield is p/n E4440-00055. This shield is only used on the top
switch.
1. Position the standoffs over the bottom switch mounting holes.
2. Reconnect W84, W86 and W87 by matching the cable information
3. Refer to
4. Attach the adhesive EMI shield to the top of the switch.
Chapter 11
Position the replacement top switch with the label up, and install
the mounting screws.
bands to the switch connector designators. Torque to 10 in-lbs.
Figure
11-19. Re-solder the wires removed earlier.
shows the proper position of the shield. Assure the switch port
numbers are not covered.
Assembly Replacement Procedures
RF Section E4440A, E4443A, E4445A
Figure 11-20
333