Post-Repair Procedures

Post-Repair Procedures
Post-Repair Procedures
Table 12-1
reset procedures, and performance verification tests needed after an
assembly replacement. Adjustments and performance verification tests
and the utility that resets the calibration constants and resets the
statistical data are located in the Performance Verification and
Adjustment Software.
Calibration constants are stored on most board assemblies. Several
PSA model numbers share the same board assembly, but require
different calibration constants.
After an assembly is replaced, find the assembly that has been replaced
in the left-hand column, and then perform the adjustment, test, or
configuration procedure shown in the adjustment column. After
successfully completing the necessary adjustment, test, or configuration
procedure, verify that the instrument meets specifications by running
the performance verification test or tests listed.
Refer to your user's guide for information on instrument warm-up
NOTE
before performing any of the procedures listed in this chapter.
Table 12-1
Post-Repair Testing Requirements
Assembly
A1 Flat Panel Display
A2 Front Panel
Interface Assembly
A3 Keyboard Assembly
A5 Power Supply
A6 SCSI Board
412
lists the adjustments, configuration or calibration constant
Adjustments, Tests, &
Configuration
Ensure that the display quality is
satisfactory. There should be no light
or dark lines or pixel "triplets".
Perform the Internal Front Panel test.
Press
,
,
System
More
Diagnostics
Ensure that the display quality and
brightness are satisfactory.
Ensure that the green "power on" and
yellow 'standby" LEDs function
properly.
Internal Front Panel test
Verify that the instrument powers up
correctly.
Check all power supply voltages using
an extender board.
None
Performance Verification
None
None
.
None
None
None