Removal - Agilent Technologies PSA Series Service Manual

Spectrum analyzer
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Removal

1. Using a 9/16" socket drive, remove the nut securing the
2. Refer to
3. Lift the front frame assembly away from the deck.
Figure 11-5
Front Frame Assembly Removal
Chapter 11
BNC connector. Take care to not scratch the front dress panel.
Input
Figure
11-5. Disconnect the ribbon cable (3) from the A2
front panel interface board. Pull the coaxial cable (4) from the front
frame and unclip from the two cable clamps.
Assembly Replacement Procedures
Front Frame
Ext Trigger