Assembly Replacement Procedures
RF Section E4440A, E4443A, E4445A
Option 123 Switches
Figure 11-19
Option 123 Switch Wiring
Top Switch Removal
1. Drop the front frame. See
2. Note the locations of the soldered wires on the switch terminals,
3. Unsolder and remove the wires from all top switch solder-on
4. Refer to
5. Remove the two switch mounting screws and remove the switch. Be
332
particularity the black wires.
locations.
terminals. When removing the black wires, avoid damaging the
wires since there is little service length. The colored wires usually
have extra length and you may wish to just cut them off and re-strip
the ends.
Figure
11-17. Remove cables W84, W86, and W87 from the
top switch. Note that two of the cables have identification bands that
correspond to the connector designator 1, or 2, printed on the switch.
careful that the standoffs between the two switches are not lost.
page 314
for instructions.
Figure 11-19
shows most wire
Chapter 11