Agilent Technologies PSA Series Service Manual page 205

Spectrum analyzer
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Oscilloscope Setup:
Input 1
Turn on the input.
1 M ohms,
DC Coupling,
BW limit OFF
Probe = 1
Input 2
Turn on the input.
1 M ohms,
DC Coupling,
BW limit OFF
Probe = 1
Invert = Off
Time/Div
200 ms
Volts/Div
200 mv
Trigger Source
Input 2
Trigger Mode
Normal
Trigger Level
Set so oscilloscope triggers
The oscilloscope display reveals the following:
The Video Out signal resembles the PSA display. This means the A7
Digital IF assembly in the PSA has successfully reconstructed a video
signal from the ADC.
The period of the trigger signal is 1 ms which corresponds to the full
10 division horizontal sweep time of 1 ms on the PSA.
The 200 mv /Div vertical scope setting allows 5 vertical divisions on the
oscilloscope to correspond to the 10 division display on the PSA.
Chapter 8
Hardware Options
Verifying Option 124, Y-Axis Video Out
This input connected to
PSA Video Out
This input connected to
Trigger 1 Out of PSA.
Displaying the external
trigger signal shows the
relationship between the
trigger signal and the
displayed video on the
oscilloscope.
205