Assembly Replacement Procedures
Rear Frame
Rear Frame
Use ESD precautions when performing this replacement procedure.
CAUTION
Removal
1. Remove the instrument top brace. Refer to the
2. Refer to
3. Remove the vertical board assemblies. Refer to the
4. Remove the A6 SCSI board. Refer to the
5. Remove the A26 CPU assembly. Refer to the
6. Refer to
7. Refer to
390
procedure.
Figure
11-51. Unplug the W12, W13, W14, W19, W21, and
W22 cables from the boards. If the instrument has optional
hardware, there will be additional cables to remove.
Assemblies (Standard Instrument)"
procedure.
removal procedure.
Figure
11-52. Remove the rear dress panel, which includes
numerous cables. Using the T-10 driver, remove the 13 screws (1).
Figure
11-53. Using a T-10 driver, remove the 11 screws (1)
that secure the rear frame to the deck. Remove the rear frame from
the deck.
"Top Brace"
"Vertical Board
removal procedure.
"A6 SCSI Board"
"A26 CPU Assembly"
removal
removal
Chapter 11