Agilent Technologies PSA Series Service Manual page 182

Spectrum analyzer
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Hardware Options
Verifying Option 122 or 140, Wide Bandwidth Digitizer
11.Press
and turn the dither signal
and
to view
WB ADC Dither
On
Off
the effect of turning dither on and off.
12.Press
Analog Filter
13.Select
Narrow 36 MHz
The PSA display now shows the filter response of the 36 MHz filter
path in the Wideband Analog IF assembly. See
Figure
8-7.
Figure 8-7
36 MHz Filter Response
If the response of only one of the filter paths is incorrect, the problem is
isolated to that filter path on the A31, Wideband Analog IF assembly.
182
Chapter 8