Improper displayed signal amplitudes in some attenuator settings can
be isolated to either A13 or A14 based on which steps are incorrect.
Signal out of the attenuators can be measured by disconnecting the
appropriate semirigid cable and using a spectrum analyzer. Referring
to
10 dB, a 20 dB and a 30 dB section. Verify that the attenuators are
receiving the correct switching sequences by using the following tables.
Table 4-1
A15 Settings
Attenuation
30 dB
(dB)
Select J4
pin 9
0
High
10
High
20
High
30
Low
40
Low
50
Low
60
Low
70
Low
Table 4-2
A14 Settings
Attenuation
2 dB - B
(dB)
Select
J3 pin 9
0
High
2
High
4
Low
6
High
8
High
10
High
Chapter 4
Figure
4-4, A14 has two 2 dB sections and one 6dB section. A15 has a
30 dB
Bypass J4
Select J4
pin 4
Low
Low
Low
High
High
High
High
High
2 dB - B
Bypass
J3 pin 4
Low
Low
High
Low
Low
Low
Troubleshooting the RF Section (E4446A, E4447A, E4448A)
RF Section Description (E4446A, E4447A, E4448A)
10 dB
10 dB
Bypass J4
pin 2
pin 1
High
Low
Low
High
High
Low
High
Low
Low
High
High
Low
Low
High
Low
High
2 dB - A
2 dB - A
Select
Bypass
J3 pin 8
J3 pin 5
High
Low
Low
High
Low
High
High
Low
Low
High
High
Low
20 dB
20 dB
Select J4
Bypass J4
pin 8
pin 5
High
Low
High
Low
Low
High
High
Low
High
Low
Low
High
Low
High
Low
High
6 dB
6 dB
Select
Bypass
J3 pin 7
J3 pin 6
High
Low
High
Low
High
Low
Low
High
Low
High
High
Low
97