Agilent Technologies PSA Series Service Manual page 399

Spectrum analyzer
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Display Removal
Work in a clean environment to avoid getting dust on the display. The
CAUTION
new flat panel display comes with a protective plastic sheet over the
glass. Remove this plastic very slowly to avoid damage due to ESD.
The surface of the display is very easily scratched. Avoid touching it
CAUTION
with your bare hands or other objects. Use a blower to remove any dust
from the display surface.
1. Refer to
2. Unplug the inverter backlight cable from the top inverter board and
3. Unplug the bottom backlight cable at the backlight extension cable
4. Remove the display mount / display from the front frame.
5. Remove the pressure plate (8) from between the display mount at
6. Peel back the rubber tabs on the display mount and carefully remove
Chapter 11
Figure
11-59. Unplug the Display Converter Board (10)
from the Flat Panel Display. Peel back the attached flat flex cable
(3) with the tape still attached until the ferrite block is free of the
rubber mount.
pull cable free of the rubber display mount (5).
(4) and pull the backlight extension cable free of the rubber display
mount.
the top of the display.
the display.
Assembly Replacement Procedures
Front Frame Subassemblies
399