Agilent Technologies PSA Series Service Manual page 99

Spectrum analyzer
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Verifying the A21 FELOMA Assembly
First check that the YTO signal is present at the LO IN connector by
measuring the power at the cable from the YTO. The signal should be
3-7 GHZ at +12 to +17 dBm.
Next verify the signals at the LO outputs. The Lowband, RYTHM, LO
Out and Sampler outputs all range from 3 GHz to 7 GHz. The SBTX
output is doubled, and will range from 6 to 14 GHz.
Table 4-3
To verify the SBTX output at J2, set the PSA to a center frequency
above 26.8 GHz and set the span to zero Hz.
Verify the RYTHM output at J3 by setting the center frequency of the
PSA from 3 GHz to 26.5 GHz while in zero span. Notice the table
includes some approximate out of band measurements, shown in italics,
that can be useful if you suspect a port-to-port isolation problem.
Verify the Lowband output at J4 by setting the center frequency of the
PSA from 3 Hz to 2.85 GHz while in zero span. Again, additional out of
band values are given for isolation purposes.
Verify the Sampler output and the 1ST LO outputs while in zero span.
Currently the J6 LO Output is not leveled.
Table 4-3
Example measurements for FELOMA outputs
PSA
J2
Center
SBTX
Frequency
Power/Fre
q
50 MHz
NA
10 GHz
NA
+13 dBm
30 GHz
13.040 GHz
Chapter 4
Troubleshooting the RF Section (E4446A, E4447A, E4448A)
lists the LO power at the different FELOMA output ports.
J3
J4
RYTHM
Lowband
Power/Freq
Power/Freq
−21 dBm
+14 dBm
3.971 GHz
3.9714 GHz
(typical)
+13 dBm
−25 dBm
5.161 GHz
5.161 GHz
(typical)
−14 dBm
−17 dBm
6.519 GHz
6.519 GHz
(typical)
(typical)
RF Section Description (E4446A, E4447A, E4448A)
J5
Sampler
Power/Freq
−14 dBm
3.971 GHz
−9 dBm
5.161 GHz
−11 dBm
6.519 GHz
J6
st
1
LO Out
Power/Freq
+14 dBm
3.971 GHz
+13 dBm
5.161 GHz
+15.95 dBm
6.519 GHz
99

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4440aE4443aE4445aE4446aE4447aE4448a

Table of Contents