Trigger Out Smb Port; Break In Smb Port; Power Supply; Electrical Characteristics - Agilent Technologies E5900B User Manual

Hide thumbs Also See for E5900B:
Table of Contents

Advertisement

Specifications and Characteristics
Input/Output Electrical Characteristics
Input/Output Electrical Characteristics

Trigger Out SMB Port

With a 50
load, a logic high is >= 2.0 V, and a low is <= 0.4 V. The
9
output function is selectable (see "To configure Trigger Out" on
page 69).

Break In SMB Port

Edge-triggered TTL level input, 20 pF, with 4.6 k
to ground in
9
parallel. Maximum input: +5 V to -5 V when the trace port analyzer is
powered OFF; +10 V to -5 V when the trace port analyzer is powered
ON. Input function is selectable (see "To configure Break In" on
page 69).
Communication Ports
Serial Port
RJ12 connector (DB9-to-RJ12 adapter and serial cable included). RS-
232 DCE to 9600 baud, 8-bit, no parity, one stop bit.
IEEE 802.3 Type 10/100Base-TX LAN Port
RJ-45 connector, is compatible with both 10 Mbps (10Base-T) and
100 Mbps (100Base-TX) twisted-pair ethernet LANs.

Power Supply

Input. 100-240 V, 1.0 A, 50/60 Hz, IEC 320 connector.
Output. 12 V, 3.3 A
CAT I (Mains isolated). Pollution degree 2.
Emulation for the Motorola M•CORE
117