A6 Power Supply Assembly Replacement; How To Remove The Power Supply Assembly - Agilent Technologies 8711C Service Manual

Rf network analyzers and tuner analyzer
Table of Contents

Advertisement

A6 Power Supply Assembly Replacement

he power supply is a single assembly. he only lower level part available for replacement is
the cooling fan.
How to emove the Power Supply Assembly
1. Disconnect the power cord.
2. Remove the front panel assembly (see \How to Remove the Front Panel Assembly").
Remove the handle nut plate on the left side only.
3. Remove the display enclosure (see the following section \How to Remove the Display
Assembly").
4. Remove the four screws along the rear edge of the display enclosure.
5. Remove the screw from the display access panel.
6. Open the display access panel and disconnect the two-wire cable from the power supply.
7. Withdraw the power supply from the housing.
3-12
Assembly Replacement

Advertisement

Table of Contents
loading

This manual is also suitable for:

87128713c8714c8730a