Comparator Function On/Off; Auxiliary Bin [Aux] - Teledyne T3LCR1002 User Manual

Precision lcr meters
Hide thumbs Also See for T3LCR1002:
Table of Contents

Advertisement

1.800.561.8187
This page provides the following information:
FUNC, RANGE, FREQ, LEVEL, TRIG, LEVEL,
SPEED: These conditions can be set from
[MEAS DISPLAY] page.
Comparator function ON/OFF [COMP].
Auxiliary bin ON/OFF [AUX].
T3LCR built-in comparator can sort DUTs into a
maximum of 10 bins (BIN1 to BIN9 and OUT) using a
maximum of nine pairs of primary parameter limits and
one pair of secondary parameter limits.
A DUT which has the primary parameter within the
limits, but secondary parameter is not, can be sorted into
an auxiliary BIN (AUX).
See Figure 5-3 on page 57 for the Comparator Workflow.
Step 1.
Press the [Setup] key.
Step 2.
Press the [BIN MEAS] soft-key.
Step 3.
Use the cursor key to select [COMP] field.
Step 4.
Use the soft-keys to turn on/off the
comparator.
www.
.com
T3LCR Series User Manual
63
information@itm.com