Teledyne T3LCR1002 User Manual page 38

Precision lcr meters
Hide thumbs Also See for T3LCR1002:
Table of Contents

Advertisement

36
1.800.561.8187
OFF
Disables open correction.
MEAS OPEN
Starts open correction (AC).
DCR OPEN
Starts open correction (DC).
Step 4.
Press the [MEAS OPEN] or [DCR OPEN]
soft-key. A dialog message,
Step 5.
Connect the test fixture to the BNC terminals.
Do not connect the DUT to the fixture.
Step 6.
Press the [OK] soft-key. The T3LCR will then
measure the open admittances (capacitance
and conductance) at the all test frequency
points.
shown on the display.
When the measurement has finished,
measurement, you can press the [ABORT]
soft-key to abort the OPEN correction.
Step 7.
After the OPEN measurement has finished,
the meter will enable OPEN correction
automatically. You can, however, still choose
to manually enable or disable OPEN
correction by depressing the ON or OFF soft-
keys.
The short correction feature of the T3LCR Series
compensates for any residual impedances (R,X) that may
exist outside of calibration, such as the impedance of the
cables and the DUT connection points.
See Figure 4-4 Residual Impedances.