Step 1.
Press the [MEAS SHORT] soft key, a
dialog message,
, will be displayed.
Step 2.
Connect the test fixture to the BNC terminals
and short-circuit the test terminals.
Step 3.
Press the [OK] soft-key. The T3LCR measures
the short impedances (inductances and
resistance) at this frequency point.
m
be shown on the display.
When the measurement has finished,
screen. During the measurement, you can press
the [ABORT] soft-key to abort short
correction.
Press the [Measure] key and press the [LIST SETUP]
soft-key to open the [LIST SETUP] page. The List
Measure feature can perform automatic sweep
measurements by sweeping the frequency or signal level
through a maximum 10 list points.
Before using the List Measure feature, you have to
configure the list setup.
In the [LIST SETUP] page, you can configure each of
the following list measurement controls with the cursor
placed in the corresponding field.
Function mode [FUNC]
Test mode [MODE]
Parameter selection [FREQ[Hz], VOLT[V],
CURR[A]]
Limit parameter selection [LMT]
Lower and upper limits [LOWER] [UPPER]