Teledyne T3LCR1002 User Manual page 46

Precision lcr meters
Hide thumbs Also See for T3LCR1002:
Table of Contents

Advertisement

44
1.800.561.8187
Trigger mode
When entering LIST MEAS for the first time, the system will
set Trig Mode to MAN.
Trig Mode
Function
INT
Internal Trigger. All ten list points are
swept continuous.
MAN
Manual Trigger. Each time the instrument
is triggered by [Trig]
key, the list points are swept one by one.
EXT
External Trigger. Each time the
instrument is triggered by the
handler trigger pin, the list points are
swept one by one.
BUS
BUS Trigger. Each time the instrument is
triggered by SCPI command, the list
points are swept one by one.
The [LIST MEAS] page will execute a list of up to 10
swept frequency or amplitude tests.
When [MODE] is set to SEQ and [TRIG] is set to
MAN, the [LIST MEAS] function will automatically
execute each test step on the list in sequence until the
last step is executed. The meter will then wait for the
trigger button to be depressed before repeating the
process.
When [MODE] is set to STEP and [TRIG] is set to
MAN, the [LIST MEAS] function will automatically
execute the first test step. The meter will then wait for
the trigger button to be depressed before executing the
next step. This process is repeated for each step on the
list.
Test Mode
Soft-key
Function
SEQ
Sequence measure all step by one
trigger.
STEP
Measure one step by one trigger.
www.
.com
Measure Key
information@itm.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

T3lcr1100T3lcr1300Everywhereyoulook t3lcr1002Everywhereyoulook t3lcr1100Everywhereyoulook t3lcr1300

Table of Contents