Before saving the recorded readings in the internal buffer
to the external USB flash drive, remember to plug the
external USB drive into the USB port on the front panel.
The recorded readings will be saved in a subdirectory
named MEAS DATA, whose parent directory will have
the same name as the T3LCR model being used. See the
example in below:
Example:
F:\T3LCR1300\MEAS DATE\ MEAS0000.CSV
Up to 9999 recording files can be saved, with their file
names from 0001.csv-9999.csv.
The size of the internal buffer can be adjusted in
setting field. The maximum data buffer size is 10000.
When you press the [Measure] key and the [OPEN
SHORT] soft-key, the [ OPEN SHORT ] page appears.
In this page, the OPEN/SHORT/SPOT correction for
correcting the stray admittance and residual impedances
can be performed.
The OPEN and SHORT function performs open or
short correction over a range of fixed trimming points.
The correction for all other frequencies are interpolated
from those fixed trimming points. Please refer to the 12.1
Specification page to see what the trimming frequency
points are for each T3LCR model.
The SPOT function will perform open, or short
correction at a user-
In the [ OPEN SHORT] page, you can configure each of
the following controls with the cursor placed in the
corresponding field:
Open correction [OPEN]

Short correction [SHORT]

SPOT correction [SPOT]