86
1.800.561.8187
Step 1.
Turn the T3LCR Series ON.
Step 2.
Set up the measurement conditions by filling in
the fields on the MEAS DISPLAY page.
Move to the FUNC field using the cursor
keys and choose Cs-D.
Move to the FREQ field using the cursor
keys and input 1 kHz.
Move to the LEVEL field using the cursor
keys and input 1 V
Step 3.
Connect the test fixture to the T3LCR Series.
Step 4.
To compensate for the test fixture, a user
correction is required.
below:
Press the [Measure] key and the [OPEN
1.
SHORT] soft-key.
Move to the OPEN TEST or SPOT field
2.
using the cursor keys.
3.
shown here:
Press the [MEAS OPEN] soft-key and
4.
then the [OK] soft-key. Wait until the
displayed.
Press the [ON] soft-key if the OPEN field
5.
is set to OFF.
Move to the SHORT TEST or SPOT
6.
www.
.com
the unit, as shown
connect the clips to anything, as
message is