Soft-key
Function
30
50
100
The averaging feature allows you to obtain moving
average values of successive measurement results.
You can specify the averaging factor from integer 1 to
integer 256.
Step 1.
Press the [Setup] key.
Step 2.
Use the cursor key to select [AVG] field.
Step 3.
Use the soft-keys or number keys to enter
averaging factor.
Soft-key
Function
INCR +
Increments the averaging factor in
steps of 1, 2, 4, 8, 16, 32,
64, 128 and 256.
DECR -
Decrements the averaging factor in
steps of 1, 2, 4, 8, 16, 32,
64, 128 and 256.
This function will apply a DC bias voltage onto the DUT
while LCR meter is performing its AC measurement on
the DUT. The settable DC bias voltage range is from -
2.5V-2.5V.
Step 1.
Press the [Setup] key.
Step 2.
Use the cursor key to select [BIAS] field.
Step 3.
Use the soft-keys to select DC bias voltage.
Soft-key Function
OFF
Turn off the DC bias voltage
2V
2V voltage is applied to the device under test.
1.5V
1.5V voltage is applied to the device under test.
-1.5V
-1.5V voltage is applied to the device under test.
-2V
-2V voltage is applied to the device under test.
www.
.com
T3LCR Series User Manual
51