Step 1.
Press the [Measure] key.
Step 2.
Press the [OPEN SHORT] soft-key.
Step 3.
Use the cursor key to select the [SHORT] field.
Soft-key
Function
ON
Enables short correction.
OFF
Disables short correction.
MEAS SHORT
Starts short correction (AC).
DCR SHORT
Starts short correction (DC).
Step 4.
Press the [MEAS SHORT] or [DCR SHORT]
soft-key, a dialog message,
Step 5.
Connect the test fixture to the BNC terminals
and short-circuit the test terminals.
Step 6.
Press the [OK] soft-key. The T3LCR measures
the short impedances (inductances and
resistance) at the all test frequency points.
shown on the display.
When the measurement has finished,
measurement, you can press the [ABORT]
soft-key to abort the SHORT correction.
Step 7.
After the SHORT measurement has finished,
the meter will enable SHORT correction
automatically. You can, however, still choose
to manually enable or disable SHORT
correction by depressing the ON or OFF soft-
keys.
The spot correction function involves performing
open/short correction at user-specified frequency points.
You can specify 1 frequency point.