Teledyne T3LCR1002 User Manual page 75

Precision lcr meters
Hide thumbs Also See for T3LCR1002:
Table of Contents

Advertisement

1.800.561.8187
If the Result setting it set to Auto, the meter will
automatically send out the measurement results each time
a test is finished; this setting is convenient especially
when the meter is working with a sorting machine. The
meter will start a test after receiving the trigger signal and
then returns the test result to the sorting machine
without the need to receive
either the sorting machine or the control PC.
Step 1.
Press the [Measure] or [Setup] key.
Step 2.
Press the [SYSTEM] bottom soft-key.
Step 3.
Use the cursor key to select the [RESULT]
field.
Step 4.
Use the soft-keys to set.
Soft-key
Function
FETCH
The meter will not send out the test
result after the test.
AUTO
The meter will send out the test result
after the test.
The data buffer setting sets the maximum number of
record counts that the internal buffers can store; Settable
range 1-10000.
Step 1.
Press the [Measure] or [Setup] key.
Step 2.
Press the [SYSTEM] bottom soft-key.
Step 3.
Use the cursor key to select the
[DATA BUFFER] field.
Step 4.
Enter the number of sets using the entry keys
or use the soft-key to select the maximum
number of sets.
Soft-key
Function
MAX
Sets the maximum record counts to be
10000 directly.
www.
.com
T3LCR Series User Manual
73