Step 1.
Press the [Setup] key.
Step 2.
Use the cursor key to select [MON 1] or
[MON 2] field.
Step 3.
Use the soft-keys to select DC bias voltage.
Soft-key
Function
OFF
Turn off the monitor
Z
Absolute value of impedance
D
Dissipation factor
Q
Quality factor(=1/D)
Vac
Test signal Voltage
Iac
Test signal Current
Absolute deviation value
Relative deviation value
r
Phase radian
d
Phase angle
R
Resistance (=Rs)
X
Reactance
G
Conductance
B
Sustenance
Y
Absolute value of admittance
This setting decides how long the meter will wait to
commence a measurement after the trigger signal is
received. Settable range: 0ms to 60s.
The automatic level control (ALC) feature adjusts the
voltage across the DUT or the current through the DUT
to match the voltage/current level setting. Using this
feature, you can try to ensure a constant signal level
(voltage or current) is applied to the DUT.